Calibration of natural gammaray spectrometer facility at the ERL
Authors:
Khodja Mohamed R Aksoy Abdulkadir
Issue Date:
1999
Citation:
Calibration of natural gammaray spectrometer facility at the ERL Abdulkadir Aksoy and Mohammad R KhodjaThe Arabian journal for science and engineering المجلة العربية للعلوم والهندسة Univeristy of Petroleum and MineralsVol 24 no 1A (January 1999) p p 7379Aksoy AbdulkadirKhodja Mohamed R
Abstract:
Gammaray spectrometer facility at the Energy Research Laboratory (ERL) is used for
the determination of natural radioactivity from potassium (40K) uranium (238U) and thorium
(32Th) in soil core and various other industrial samples The facility consists of a
S x S NaI (Tl) detector and the electronic setup coupled to a PCbased data acquisition
and analysis system The gammaray spectrometer facility was calibrated using 12 certified
standard samples from IAEA (Vienna) NIST and NBL (U S A) The sensitivity of the
setup was found to be better than 1 ppm for U and Th and 150 ppm for K considering
100 g sample Linear calibration curves were established from measurements carried out
for potassium uranium and thorium يستخدم مطياف أشعة جاما الموجود في مختبر أبحاث الطاقة لتحديد وقياس النشاط الإشعاعي الطبيعي الصادر من البوتاسيوم (K 40) واليورانيوم (U 238) والثوريوم (Th 238) في التربة وعينات صناعية مختلفة ويتكون المطياف من كاشف (Tl)NaI 5 × 5 وتجهيزات إليكترونية موصلة بحاسوب لتجميع وتحليل النتائج التجريبية ولقد تم معايرة مطياف أشعة جاما باستخدام اثنتا عشر عينة قياسية معتمدة من (IAEA) في فينا و(NIST) و(NBL) في أمريكا ولقد وجد بأن حساسية هذا النظام تصل إلى جزء من المليون لعنصر اليورانيوم وعنصر الثوريوم و150 جزءا من المليون لعنصر البوتاسيوم على اعتبار أن وزن العينة 100 غم كما أنه تم تعيين منحنيات المعايرة الخطية لهذا النظام من خلال عمل بعض القياسات لعناصر البوتاسيوم واليورانيوم والثوريوم