A high frequency oscillator based auto titrimetric method for redox systems
المؤلفون:
Jaffar M
تاريخ النشر:
1991
الاستشهاد المرجعي :
A high frequency oscillator based auto titrimetric method for redox systems M Jaffar [et al]The Arabian journal for science and engineering المجلة العربية للعلوم والهندسة Univeristy of Petroleum and MineralsVol 16 no 2A (April 1991) p p 153158Jaffar M
الملخص:
An autooscillometric technique has been developed for tracing the equivalence points in aqueous redox estimations as a function of titrant volume The technique utilizes the current impulses generated in a transistorized circuit oscillating at 110 MHz for the autorecording of the titration events The usefulness and application of the technique is demonstrated by selecting various reductionoxidation systems in aqueous phase at concentrations ranging between 103 to 106 M The overall error is rated at about ± 11 % with a precision better than ± 1 % for replicate measurements on a given redox system This technique has the potential of application for analytical estimations related to compleximetry and permittivity changes طور أسلوب (الأوسيلوميتر) الأوتوماتيكي لتحديد نقاط التعادل في محاليل (ريدوكس) المائية كدالة للمحلول المعاير ويستغل هذا الأسلوب الذبذبات التيارية في دائرة كهربائية من الترانزستور تتذبذب في (110) ميغا هيرتز للتسجيل الأتوماتيكي لمختلف عمليات المعايرة وأوضحت الطريقة إفادة وتطبيق هذا الأسلوب باختيار مختلف أنظمة الأكسدة والختزال في محيط مائي في تراكيز تتفاوت فيما بين (103) إلى (106) مول أما مجمل الخطأ فإنه يقدر بحوالي (± 11 %) وبدقة أحسن من (± 1 %) في القياسات المماثلة ريدوكس (أكسدة اختزال) ويمكن أن يستخدم هذا الأسلوب في التحليلات التقديرية للتغيرات في طريقة معايرة محاليل المركبات المعقدة وفي تغييرات المنفذية الكهربائية